4. The elastomer rubber light shield and Tacking Tips can be
cleaned, sterilized and reused. Prior to re-use inspect for
damage and discard if damage is detected.
Sterilizing:
Package in a sterilization wrap. Sterilization can be performed
with either of the following cycles;
1. In a gravity autoclave at 132°C / 270°F for 15 minutes
with a 15 – 30 minute drying time.
2. In a pre-vacuum sterilizer at 132°C / 270°F for 4 minutes
with a 20 – 30 minute drying time.
Reassembly & Storage
1. After all parts have been allowed to dry completely,
gently insert the metal end of the glass fiber optic light
guide into the circular opening of the handpiece. Make
sure the light guide is completely inserted until it stops.
This will position the light guide properly with the internal
LED light source.
2. Slip the entire assembly into a new undamaged
polyethylene film barrier sleeve. Be careful not to cause
damage to the barrier sleeve film.
3. Fold the end of the barrier sleeve film over the tip of the
fiber optic light guide making sure the seam in the barrier
sleeve does not pass across the light exit face of the light
guide. This will allow optimum light output.
4. Slide an eye shield onto the light guide. This will hold the
polyethylene barrier sleeve taut over the tip of the light
guide and hold it in place ready for the next use.
5. Verify the polyethylene film barrier sleeve is still intact and
has not suffered damage such as tears or cuts. Store the
reassembled S.P.E.C. 3
location, preferably on the S.P.E.C. 3
so that the battery will be completely charged prior to the
next use.
®
LED Curing Light in a clean dry
®
LED charger base
EN
VIII. Technical Details
8.1. Power Supply
A. Input: 100V – 240V AC / 50 ~ 60 Hz
B. Output: S.P.E.C. 3
8.2. Dimensions
Component
Handpiece
Power Adapter
Charger Base
AC Power Cord
8.3. Environment
A. Operating Conditions
Temperature:
Relative Humidity:
Ambient Pressure:
B.
Transportation and Storage Conditions
Temperature:
Relative Humidity:
Ambient Pressure:
®
: 6V DC, 2.5A
S.P.E.C. 3
®
Curing Light
174.5 × 24.8 (mm)
47 × 87 × 32 (mm),
136 × 86 × 60 (mm),
1.3M
-5°C ~ +40°C (23°F ~ 104°F)
10% ~ 85%
80 ~ 106 kPa
(23.62 inHg ~ 31.30 inHg)
-10°C ~ +45°C (14°F ~ 113°F)
10% ~ 90%
60 ~ 106 kPa
(17.72 inHg ~ 31.30 inHg)