7.3

DC Accuracy

The gain of the probe-oscilloscope system and the accuracy of the integrated DC
voltmeter are checked.
Test setup:
R&S RTO settings:
User Manual 1410.5363.02 ─ 03
Probe connected to CH1 of the R&S RTO oscilloscope
INPUT HI/LO of the DMM connected to OUTPUT HI/LO of the DC
voltage source by banana leads
BNC to 4 mm dual banana adapter plugged onto the banana leads
on the DMM – ensure that BNC shield is connected to LO
Signal and ground terminal of the probe connected by mini clips,
leads, and socket adapter to the BNC to 4 mm dual banana
adapter; make sure that probe ground is connected to BNC shield
Horizontal scale = 10 µs / div
Vertical scale = 1.6 V / div
Trigger source = Ch1, Trigger mode = Auto, Run cont
ACQUISITION key > Decimation = 'High res' > Waveform Arithmet-
ics = 'Average' > Average Count = 16
"Meas" menu > Setup > Source = 'Ch1' > Main Measurement =
'Mean' > State = 'on'
CH1 key > Offset = 0 V, Position = 0 div
"Vertical" menu > Probe Setup > Ch1 > ProbeMeter = 'on'
Functional Check
DC Accuracy