KERN FES 17K-4 Betriebsanleitung Seite 205

Präzisionswaage
Vorschau ausblenden Andere Handbücher für FES 17K-4:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ERR751
ERR752
ERR753
ERR760
ERR761
ERR764
TFES_TFEJ-BA-def-2210
Unit weight of samples is less than the
smallest unit weight
Piece weight of samples is too low (0 g
or less) during piece counting mode
Timeout during import of piece weight
in piece counting mode
Attempt to perform totalizing operation
even though totalizing function is
disabled
Error at <636 REF CAL>
External weight for <631 EX CAL> is
different from the weight range set in
<SELECT WEIGHT>.
Use samples whose unit
weight is greater than the
smallest unit weight
Use samples whose unit
weight is greater than the
smallest unit weight
Weighing plate or weighing
plate carrier incorrectly
adjusted
Check whether the scale is
touching other objects
Check for vibrations or
draughts
Set <141 ACTIVATE> to
ON to use the totalizing
function.
Carry out <636 REF CAL>
again
Use external weight whose
weight value is in the set
range
101