ERR751
•
ERR752
ERR753
•
ERR760
•
ERR761
•
ERR764
•
TFES_TFEJ-BA-def-2210
Unit weight of samples is less than the
smallest unit weight
Piece weight of samples is too low (0 g
or less) during piece counting mode
Timeout during import of piece weight
in piece counting mode
Attempt to perform totalizing operation
even though totalizing function is
disabled
Error at <636 REF CAL>
External weight for <631 EX CAL> is
different from the weight range set in
<SELECT WEIGHT>.
•
Use samples whose unit
weight is greater than the
smallest unit weight
•
Use samples whose unit
weight is greater than the
smallest unit weight
•
Weighing plate or weighing
plate carrier incorrectly
adjusted
•
Check whether the scale is
touching other objects
•
Check for vibrations or
draughts
•
Set <141 ACTIVATE> to
ON to use the totalizing
function.
•
Carry out <636 REF CAL>
again
•
Use external weight whose
weight value is in the set
range
101