5 Pressure Connection To The Sensors

For measurements of over pressure (-10.0 mbar ... 350.0 mbar):
Connect plastic tube with internal dia of 4 mm to pressure port "+". Port "-" will not be used!
For measurements of under pressure (-350.0 mbar ... 0.0 mbar):
Plug the tube to pressure port "-". The measuring range covers then -350.0 to 0.0 mbar.
Note: All values are displayed now as positive values. No minus sign will be shown.
Example: it is possible to measure under pressure down to -350.0 mbar, the display shows then the value
350.0 (no minus sign).
For measurements of pressure differences:
Connect both plastic tubes with an internal dia of 4 mm to pressure port "+" and "-"; make sure to apply
higher pressure to port "+".

6 Error And System Messages

Display
Meaning
Low battery power, device will only continue
operation for a short period of time
Battery empty
Mains operation without battery: wrong voltage
No display or
Battery empty
confused
Mains operation without battery: wrong voltage
characters,
or polarity
device does
System error
not react on
keypress
Device defective
Err.1
Measured value above allowable range
Sensor defective
Err.2
Measured value below allowable range
Sensor defective
Err.4
Value is too low to be displayed, tara is set
Err.9
Measured value far out of allowable range
Err.7
System error

7 Calibration Services

Calibration certificates – DKD-certificates – other certificates:
If device should be certificated for its accuracy, it is the best solution to return it to the manufacturer.
Only the manufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy!
_______________________________________________________________________________________________________________________
GREISINGER electronic GmbH
User's Manual GMH3181-07
Hans-Sachs-Straße 26
D-93128 Regenstauf
What to do?
Replace battery
Replace battery
Check power supply, replace it when
necessary
Replace battery
Check power supply, replace it when
necessary
Disconnect battery and power supplies,
wait shortly, then reconnect
Return to manufacturer for repair
Check: pressure above 350 mbar?
-> measuring value to high
Return to manufacturer for repair
Check: pressure below –10 mbar?
-> measuring value to low
Return to manufacturer for repair
Check: display below -2000 (tara?)?
Check: pressure not within sensor range?
Return to manufacturer for repair
Fon: 0049-9402 / 9383-0
Fax: 0049-9402 / 9383-33
page 10 of 12