Volume level Vo1 flashes on the display.
• You can set the volume of the selected language using the func-
tion buttons
Vo3 = loud
Vo2 = medium
Vo1 = quiet
Confirm your selection with the Memory button .
Operation with the mains part
You can also operate this device with a mains part.
When doing so, there must not be any batteries in the battery compartment. The mains part can be ob-
tained from specialist retailers or from the service address using order number 071.60.
• To prevent possible damage to the device, the blood pressure monitor must only be used with the mains
part described here.
• Insert the mains part into the connection provided for this purpose on the right-hand side of the blood
pressure monitor. The mains part must only be connected to the mains voltage that is specified on the
type plate.
• Then insert the mains plug of the mains part into the mains socket.
• After using the blood pressure monitor, unplug the mains part from the mains socket first and then dis-
connect it from the blood pressure monitor. As soon as you unplug the mains part, the blood pressure
monitor loses the date and time setting but the saved measurements are retained.

5. Measuring blood pressure

Please ensure the unit is at room temperature before measuring.
Attach the cuff
Place the cuff on to the
bare left upper arm. The
circulation of the arm
must not be hindered
by tight clothing or
similar.
Caution: The unit may only be operated with the original cuff. The cuff is suitable for an arm circumference
of 22 to 36 cm.
A larger cuff for upper-arm circumferences of 35 to 44 cm can be obtained from specialist retailers or from
the service address using order number 163.387.
:
The cuff must be placed
on the upper arm so
that the bottom edge
is positioned 2 – 3 cm
above the elbow and
over the artery. The line
points to the centre of
the palm.
Now tighten the free end
of the cuff, but make
sure that it is not too
tight around the arm
and close the Velcro
fastener. The cuff should
be fastened so that two
fingers fit under the cuff.
47
Voice output
Now insert the cuff line
into the connection for
the cuff connector.

Quicklinks ausblenden: