6.Frozen milk can be gradually warmed in warm water or placed in a baby bottle warmer.
7.NEVER WARM YOUR BREAST IN THE MICROWAVE, BECAUSE YOU CAN DESTROY VITAL NUTRIENTS AND ANTIBODIES
THIS WAY. Also, uneven heating can lead to too hot areas of breast milk, which can cause burns to your baby's upper
digestive tract.
8.Never place a frozen bottle of milk in boiling water, as this may cause the bottle to burst / break. Never pour boiling
water directly into a feeding bottle. You should allow the water to cool for about 20 minutes before filling the bottle.
X. CLEANING AND STERILIZATION OF THE PUMP
Important: NEVER CLEAN WITH WATER. You can only clean it with a soft dry cloth.
Clean and sterilize your electric pump before first use and after each use, following the instructions below.
❖ STERILIZATION:
1.Disassemble the pump. You can wipe up the pump motor (mechanism) (1) with a slightly damp cloth and then you must
dry it immediately.
2. Wash all other parts with warm soap water and rinse them well. Sterilize ONLY the following components of the pump:
Silicone cylinder (2); Baby milk container (3); Silicone valve (4); Silicone breast (nipple) shield (5). The linker (6) should not
be heated and sterilized in a microwave oven.
3. The silicone valve (4) and silicone breast (nipple) shield (5) are made of silicone and should not be sterilized for a long
time under high temperature.
❖ RECOMMENDATIONS FOR CLEANING
1.Use a steam sterilizer. Boiling the parts is not recommended. The first reason is that the product will be easily damaged
if improper sterilization is applied. Second, this method of sterilization cannot achieve full effect.
2.If you do decide to use boiling as a sterilization method, pour more water into the container so that the parts are
completely covered with water. You must not leave the product unattended during the sterilization process. Do not boil
for too long - maximum 10 minutes.
3. For breastfeeding mothers who breastfeed every 2-3 hours, we do not recommend sterilizing and cleaning after each
use. Excessive disinfection and cleaning can easily lead to premature aging of parts and shorten the life of the breast
pump. Thorough cleaning of stains and sterilization once a day is recommended.
4. Store the parts in a clean container until the next use.
IMPORTANT! If you use the breast pump as a means of storing milk, you must take care to disinfect all washable parts of
the pump, because otherwise, the milk can easily spoil, as it is a product that is difficult to store.
ATTENTION! Before use, make sure that the pump body, valve, piston and cover are dry.
XI. TROUBLESHOOTING GUIDE
1. Lack of suction:
• Check if you have assembled all parts correctly and if the silicone valve is positioned correctly.
•Ensure that the silicone shield brace is fitted securely to the pump body and a perfect seal is created.
•Check whether there are any damaged parts.
•If you can not turn the pump on, check whether the battery is exhausted. Connect the adapter to the terminal for
charging the motor. Do not use it while charging as it shortens its battery life.
2. No milk is being expressed:
• Check point 1
• Ensure the pump is correctly assembled and that suction is being created.
• Relax and try again.
•We recommend you increase the levels gradually; the most suitable level is the one, with which the suction is painless.
The following is NOT true: the stronger – the better. Too strong suction can hurt the nipple.
3. Pain in the breast area where there is expression:
• If you feel pain and no milk flows during expression, maybe your mammary gland is blocked. This case may result in
mastitis. Contact a physician promptly. The complete milk expression will have good effect on mastitis protection.
•You may be using too high level of suction. You do not need to use all the suction that the pump can generate.
•Try to adjust the proper suction and frequency, starting at the first level and gradually increasing the levels.
•Before you start using it, put a warm cloth on your breasts and massage it for about 5 minutes.
4. Flow is too slow:
The speed of milk flow differs for every woman. The higher degree and level of expression do not mean faster flow of
milk. This depends on many factors – constitution of the breast, level of possibility, even the mood.
•The more regularly you use the breast pump, the quicker and easier the flow of milk. Before you start using it, put a
warm cloth on your breasts and use the massage mode for about 5 minutes before you switch to expression mode.