Herunterladen Inhalt Inhalt Diese Seite drucken

Sony LMD-X3200MD Bedienungsanleitung Seite 23

Vorschau ausblenden Andere Handbücher für LMD-X3200MD:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 50
Guidance and manufacturer’s declaration – electromagnetic immunity
The LMD-X3200MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the LMD-X3200MD should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 Vrms
150 kHz to
IEC 61000-4-6
80 MHz
outside ISM
bands
6 Vrms
150 kHz to
80 MHz
in ISM bands
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.7 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance
level
3 Vrms
c
6 Vrms
c
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the LMD-X3200MD, including cables, than the
recommended separation distance calculated
from the equation appliance to the frequency of
the transmitter.
Recommended separation distance
d
P
= 1.2
IEC 60601-1-2: 2007
d
P
= 1.2
80 MHz to 800 MHz
d
= 2.3
P
800 MHz to 2.5 GHz
IEC 60601-1-2: 2014
d
P
= 2.0
80 MHz to 2.7 GHz
P
Where
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with following symbol:
d
is the
a
7
GB

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis