Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing plate.
How to carry out adjustments
Ensure that there are no objects on the weighing pan.
In weighing mode keep
and at the same time press
first menu item "ECF-1" is displayed.
Press
, in the PCS-window the AD
converter value appears. 0.0 g is
displayed
Press
anew. The weight display
flashes. Use the numeric keys
for entering the value of the
required adjustment weight.
Place an adjustment weight of the
corresponding size on the weighing
plate and press
adjustment process.
The balance changes then
automatically into weighing mode.
* The adjustment should be made with the recommended adjustment weight (see
chap. 1 "Technical data"). Weights of different nominal values may also be used for
adjustment but are not optimal for technical measuring.
Info about adjustment weights can be found on the Internet at: http://www.kern-
sohn.com