Product description
3.1.2 Operating fluid return
The rotary vane pump is equipped with an operating fluid return as standard. If the operating fluid col-
lected in the operating fluid separator reaches a certain level, a float valve opens and the ejected oper-
ating fluid is returned to the intake port of the vacuum pump.
3.1.3 Cooling
The rotary vane pump is cooled by thermal radiation from its surface, the air current from the two fans
and the pumped gas.
In case of thermally unfavorable ambient conditions, Pfeiffer Vacuum recommends the use of a pump
version with oil/water heat exchanger.
● Air cooling (standard)
● Oil/water heat exchanger (optional)
3.1.4 Gas ballast system
The gas ballast system supplies air to the working chamber of the rotary vane pump periodically at the
beginning of the compression phase. When pumping down vapors, this air prevents condensation within
certain limits in the rotary vane pump. The gas ballast valve affects the final pressure of the rotary vane
pump.
3.2 Identifying the product
► To ensure clear identification of the product when communicating with Pfeiffer Vacuum, always
► Observe the motor-specific data on the motor rating plate attached separately.
3.3 Product features
Pump type
Hena 202
Hena 202 K
Hena 302
Hena 302 K
Tbl. 4:
80/126
Function of the operating fluid return
The operating fluid return functions only at an operating pressure of < 800 hPa.
keep all of the information on the rating plate to hand.
● Nominal pumping speed
─ 50 Hz: 200 m
─ 60 Hz: 240 m
● with max. constant intake pressure 100 hPa
● Vacuum connection and gas outlet: G 2"
● Operating fluid return via the float valve
● Gas ballast unit with shut-off valve
● Oil/water heat exchanger (optional)
As with Hena 202, in addition with:
● intake flange DN 100 ISO-F for direct assembly of a roots pump
● Base frame
● Nominal pumping speed
─ 50 Hz: 300 m
─ 60 Hz: 360 m
● max. continuous intake pressure 100 hPa
● Vacuum connection and gas outlet: G 2"
● Operating fluid return via the float valve
● Gas ballast unit with shut-off valve
● Oil/water heat exchanger (optional)
As with Hena 302, in addition with:
● intake flange DN 100 ISO-F for direct assembly of a roots pump
● Base frame
Features of the rotary vane pumps
Characteristics
3
/h
3
/h
3
/h
3
/h