Problem
Vacuum pump does not
reach the specified final
pressure
Pumping speed of vacuum
pump too low
Loss of operating fluid
Unusual noises during oper-
ation
Tbl. 10:
Troubleshooting for rotary vane pumps
Possible causes
● Measurement result dis-
torted
● Vacuum pump or con-
nected accessory conta-
minated
● Operating fluid contami-
nated
● Operating fluid level too
low
● Leak in system
● Vacuum pump is dam-
aged
● The intake line is not
suitably dimensioned
● Exhaust pressure too
high
● Operating fluid separator
leaking
● Radial shaft seal rings
(RSSR) leaking
● Operational loss of oper-
ating fluid
● Pumping system is con-
taminated or damaged
● Motor bearing is faulty
Malfunctions
Remedy
● Check the measurement instru-
ment.
● Check the final pressure without
system connected.
● Clean the vacuum pump.
● Check the components for con-
tamination.
● Operate the vacuum pump with
the gas ballast valve open over a
longer period.
● Change the operating fluid.
● Top up the operating fluid.
● Locate and eliminate the leak.
● Contact
Pfeiffer Vacuum
● Make sure that connections are
short and cross sections ade-
quately dimensioned.
● Check exhaust line outlet open-
ings and exhaust side accesso-
ries.
● Check for leaks.
● Replace the seal as required.
● Contact
Pfeiffer Vacuum
● Check the oil return.
● Clean and maintain the vacuum
pump.
● Contact
Pfeiffer Vacuum
● Change the motor.
● Contact
Pfeiffer Vacuum
Service.
Service.
Service.
Service.
115/130