8.3 Automatic correction of the reference weight (verifiable models only)

The balance has the possibility to correct the deviation automatically if the reference
weight differs from the previously adjusted value of the reference weight.
• Menu setting: [ F14�AQ�Qn], see chap. 6.11 Service mode
Operation
 Reset balance to zero or tare the empty
weighing container if necessary.
Set reference:
 Place a known number of pieces on the
balance as reference
 Wait for the stability display, than enter the
number of individual items via the numeric
keypad. Confirm within 5 sec:
„SAMP" is shortly displayed;
The reference weight will be determined
and displayed.
 Put on more items to be weighed, the
balance will correct the reference weight
and an acoustic signal sounds, if it has
been activated under menu item „F13 bee"
CPB-N / CPB-DM-BA-e-1831
Display
(example)
35