Feellife Air Mask II Benutzerhandbuch Seite 27

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 35
<5 % U T (>95 %
dip in U T ) for
0.5 cycle
Voltage dips,
<5 % UT(>95 %
short
dip in UT) for
interruptions
1cycle
and voltage
variations on
70 % U T (30 %
power
dip in UT) for
supply
25/30 cycles
IEC 61000-4-
11
<5 % UT(>95 %
dip in UT)for 5/6
sec
Power
frequency
(50/60 Hz)
3 A/m
IEC 61000-4-8
NOTE U T is the A.C. mains voltage prior to application of the test level.
The test of input/output lines is not applicable since they are shorter than 3.0m.
Guidance and manufacturer's declaration – electromagnetic immunity
– for device that is not LIFE-SUPPORTING .
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment
that it is used in such an environment.
Immunity
IEC60601
test
test level
<5 % U T (>95
%
dip in U T ) for
0.5 cycle
<5 % UT(>95
% dip in UT)
for
1cycle
70 % U T (30
% dip in UT)
for25/30
cycles
<5 % UT(>95
% dip in UT)
for 5/6 sec
3 A/m
Compliance
Electromagnetic environment –
level
guidance
26
Mains power quality
should be that of a typical
commercial and/or hospital
environment. If the user
of this device requires
continued operation during
power mains interruptions,
it is recommended at the
device be powered from
an uninterruptible power
supply or battery.
Power frequency magnetic
characteristic of a typical
location in a typical
commercial or hospital
environment.
the device should assure