9. Specifications

This device is not calibrated when purchased!
Regulations concerning environment of use:
Use this meter only for measurements in CAT I, CAT II and CAT III
environments (see §7).
Use this meter only in a pollution degree 2 environment (see §8).
Ideal working conditions include:
temperature: 0 °C to 40 °C (32 °F to 104 °F)
relative humidity: max. 80 %
altitude: max. 2000 m (6560 ft)
IP rating ........................................................................... IP20
voltage ........................................................................... 600 V
power supply ....................................... 3 x 1.5 V AAA/LR03 (incl.)
display ............................................................ LCD, 1999 counts
display dimensions ................................................... 24 x 37 mm
over-range .......................................................................... yes
continuity buzzer .................................................................. yes
diode test ............................................................................ yes
low-battery indication ........................................................... yes
ranging mode ......................................................... auto/manual
data hold ............................................................................ yes
backlight ............................................................................. yes
auto power-off ..................................................................... yes
dimensions ..................................................... 194 x 72 x 35 mm
weight (with battery) ......................................................... 230 g
storage environment
temperature ................................................. -20 °C to 60 °C
humidity ............................................................ < 90 % RH
test lead probe ....... CAT III 1000 V / CAT IV 600 V, 10 A; L = 84 cm
9.1 AC VOLTAGE
range
2 V
20 V
200 V
600 V
Impedance: 10 MΩ, max. input voltage: 600 V DC or AC rms
Frequency range: TRMS 40 Hz-1000 Hz
V. 03 – 05/07/2019
DCM120
Do not measure circuits that may contain voltages
resolution
0.001 V
0.01 V
0.1 V
1 V
> 600 V AC or DC
8
accuracy
± (1.0 % + 5)
± (1.2 % + 5)