Verfügbare Sprachen

The boso unit is intended for use in the electromagnetic environment specified below. The customer or the user of the boso unit should assure
that it is used in such an environment.
Immunity
test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE : U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601-
Compliance level
test level
± 6 kV
± 6 kV
contact
contact
± 8 kV
± 8 kV
air
air
± 2 kV for power
n.a.
supply lines
± 1 kV
for input/output
lines
± 1 kV
n.a.
differential mode
±2 kV
common mode
< 5% U
n.a.
T
(> 95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(> 95% dip in U
)
T
for 5 s
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30%.