Herunterladen Inhalt Inhalt Diese Seite drucken

Westfalia DT268 Bedienungsanleitung Seite 18

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Accuracy
Range
450 V
Input resistance 10 MΩ
DC Voltage Measurements
• Connect the black measuring cable to the COM input (7) and the red
measuring cable to the V/Ω input (6).
• Then set the rotary function switch (4) to the desired measuring range
(V
/ 600 V).
• Connect the measuring tips parallel to the circuitry and observe correct
polarity (Red to plus; black to minus) and read the voltage from the display.
• Is a minus symbol appears in the display, correct the polarity.
• Please make sure not to exceed 600 V.
Accuracy
Range
600 V
Input Resistance 9 MΩ
Resistance Measurements
Attention! Under no circumstances conduct resistance measurements on
voltage carrying circuitry!
• Connect the black measuring cable to the COM input (7) and the red
measuring cable to the V/Ω input (6).
• Then set the rotary function switch (4) to the desired measuring range
(Ω / 20K or 2M).
• Connect the measuring cables to the resistor you intend to test.
• Read the result from the display.
Resolution
1 V
Resolution
1 V
Operation
Accuracy (50 - 500 Hz)
± 1 % of Measured Value + 4 Digits
Accuracy
± 0.8 % of Measured Value + 1 Digit
13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis