Technische Daten
Das Alarmtelegramm hat folgenden Aufbau:
Telegrammkopf
"BEL"
Startzeichen
050
Identifikationsnummer
0
Kanalnummer
Telegrammrumpf
"ESC"
(Alarm/Warnung/Hinweis) (Status)
(Alarmnummer) (Alarmtext)
Die einzelnen Felder haben folgende Bedeutung:
01
Device failure
02
Air supply down
03
O2 supply down
04
pressure meas. inop
05
O2 measurement inop
06
flow measurement inop
07
mixer inop
08
exp. valve inop
09
fan 1 defect
10
temperature meas. inop
12
temperature high
13
flow sensor ?
14
PEEP high
15
CO2 measurement inop
16
CO2 sensor ?
17
clean CO2 cuvette
18
CO2 zero ?
22
apnoea
23
FiO2 high
24
FiO2 low
25
MV low
26
MV high
27
airway pressure low
28
airway pressure high
29
fail to cycle
30
high frequency
32
volume not constant
33
ASB > 4 s
34
etCO2 high
35
etCO2 low
240
36
air supply pressure high
37
air supply high
38
apnoea ventilation
39
Insp. hold interrupted
40
loss of data
41
Flow monitoring off
42
Monitoring FiO2 off
43
Monitoring CO2 off
44
Monitoring SpO2 off
45
O2 supply high
46
fan 2 defect
47
malfunction fan 2
48
malfunction fan 1
49
SpO2 low
50
SpO2 high
51
pulse low
52
pulse high
53
no pulse
54
SpO2 sensor ?
55
SpO2 meas. inop
57
battery not loaded
58
battery only for 2 min.
59
int. battery activated
60
ext. battery wrong
61
PEEP valve inop
62
neo. flow meas. inop
63
standby activated
64
nebulizer on
65
Tidalvolume high
67
check evita
68
frequency ILV Slave ?
69
pressure limited
70
ILV sync. inop
71
MEDIBUS inop
73
ASB > 1.5 s
74
Leakage
75
neo.flow monitoring off
76
neo.flowsensor unsuitable
77
nebulizer off
78
PPS-insp.> 1.5 s
79
PPS-insp.> 4 s
80
ASB > Tinsp
81
backup ventilation
82
Exp. hold interrupted
Gebrauchsanweisung Evita XL SW 7.0n