Operating messages – Error codes
GB
Operating messages
Hı
Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Lo
Replace batteries, no further tests possible.
btLo
Battery capacity is too low for the display backlight;
measurement is still possible.
Store
Date
A user calibrated method is indicated by a "Cal" symbol while the test
Print
Time
Cal
RESULT
result is displayed. (see "Factory calibration reset").
Error codes
Light absorption too great. Reasons: e.g. dirty optics.
E27 / E28 / E29
Calibration factor "out of range"
E 10 / E 11
Too much light reaching the detector.
E 20 /E 21
Too much light reaching the detector.
E23 / E24 / E25
E 22
Battery capacity was too low during measurement. Change battery.
E 70
Al (Tablet):
E 71
Al (Tablet):
E 72
Al (Powder Pack):
E 73
Al (Powder Pack):