6.8.

Measuring mode

Display:
The measuring mode can be selected by pressing
(C):
Continous measuring: The measured values will be displayed continously in
the display as long as the measuring key is pressed at the first step. The
actually measuring value will be written to the storage if the measuring key is
pressed at the second step.
(P):
Peak picker mode: The measured values are computed continously to an
actual maximum value. The maximum value will be written to the storage by
pressing the measuring key to the second step.
(A):
Average mode: The measured values are computed continously to an actual
average value. The average value will be written in the storage by pressing
the measuring key at the second step.
Notice:
Neither the peak picker nor the average mode compute over all values and
rows in storage but only over the measuring values got during the time the
measuring key is pressed at the first step.
6.9.

Date and time

Date and time can be selected in six steps. The input is restricted to possible values.

6.10. Factory settings

Factory settings are as follows:
Emissivity:
Minimum temperatur:
Energy saver mode:
Automatic shutdown:
Temperatur unit:
Measuring mode:
or
0.99
700
180 s
3600 s
Deg C
C
or
-
+
or
. Possible modes are:
°C
5