Measurement protocol (examples)

4.2.3

Finishing the measuring cycle

After completing the printing
process READY is displayed.
Readiness for operation is re-
established
for
the
measuring cycle.
4.2.4

Printing an additional measurement protocol

Select Print Last Test to print a copy of the last
measurement protocol at a later date.
A measurement protocol generated at a later date includes all
data of the last measurement.
NOTICE
i
i
An additional measurement protocol can only be
printed if no new measurement has been started yet
and the instrument has not been switched off.
4.3

Errors during the measuring process

If errors are detected during the measuring process, these are
indicated on the display and, dependent on the kind of error,
the provision of the breath sample must be repeated or the
measuring cycle is cancelled completely.
MINIMUM VOLUME NOT ACHIEVED
The volume of the breath sample provided is too small.
For breath samples with both too small exhalation volume and
too short exhalation time, the error message concerning the
breath volume is shown.
The provision of the breath sample has to be repeated. (A total
of 3 unsuccessful attempts per measuring cycle are
permissible.)
TEST REPEATED
Repeating a breath sample: The measuring cycle is continued
if the following message is indicated, e.g. due to an
insufficiently long delivery of the breath sample (less than
3 seconds).
CHECK AIRWAY
The controlled flow during purging is too low. Always make
sure that the air inlet into the breath hose is not covered during
purging.
TEST ABORTED
After a fourth unsuccessful attempt the measuring cycle is
terminated and a corresponding message for the cause of the
last unsuccessful attempt is issued with the protocol on the
printer.
MOUTH ALCOHOL
The presence of residual mouth alcohol was detected from the
concentration/time
profile
The measuring cycle is cancelled immediately and a protocol
printed with the respective information.
READINESS TO BLOW EXPIRED
If, after the start of the prompt, the provision of a breath sample
is not started within 3 minutes, the measuring cycle is
cancelled. A protocol with the respective information is printed.
22
next
of
the
breath
5
Measurement protocol (examples)
(Example:
measurement)
5.1
Explanations concerning the
measurement protocol
The details in the top part of the measuring protocol facilitate
the unique identification of the measuring instrument with
which the measurement was carried out and the protocol
generated.
Serial number
Unambiguous information (4 characters, 4-digit number)
assigned to only one instrument.
Sample number
Number of measurements with measurement protocol
printout after the last service at which the counter was
reset.
Calibration number
sample.
Control number which is changed when adjusting the
instrument.
The remaining details on the measurement protocol are self-
explanatory.
normal
(Example: Cancellation of the
measuring cycle)
Dräger Alcotest 9510 Standard IR / EC