Herunterladen Inhalt Inhalt Diese Seite drucken

Error Description; Error Codes - IKA LR 1000 basic Betriebsanleitung

Vorschau ausblenden Andere Handbücher für LR 1000 basic:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9
The fault is shown by an error message in the display as following if the error occurs.
Proceed as follows in such cases:
Switch off the device using the main switch
Carry out corrective measures
Restart the instrument
Error
Description
code
Error 3
Internal temperature of device is too
(Er 3)
high.
Error 4
Motor is blocked or overloaded.
(Er 4)
Error 10
Remote control is interrupted.
(Er 10)
Error 11
External temperature sensor plugged
(Er 11)
during heating control.
Error 12
External temperature sensor unplugged
(Er 12)
during heating control.
Error 14
Short circuit of external temperature
(Er 14)
sensor or cable.
Error 17
The temperature of safety sensor is
(Er 17)
higher than safety temperature setting.
Error 51
Mains voltage is too high.
(Er 51)
Error 52
Mains voltage is too low.
(Er 52)
Power is interrupted when heating or
Er 60 
motor control is activated in Mode B
and C
 Only for LR 1000 basic!
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
Contact the service department
Send the instrument for repair, including a short description of the fault.
50

Error codes

Effect
Corrective action
Heating off
- Switch off the instrument and allow it to cool
Motor off
down.
Motor off
- Switch off the instrument
- Decrease the load and restart again
Heating off
- Change watchdog time
Motor off
- Check relating connection
Heating off
- Unplug the external temperature sensor.
Heating off
- Plug in the external temperature sensor
Heating off
- Check the external temperature sensor and cable
Heating off
- Increase the safety temperature setting or allow
cooling down
Heating off
- Switch off the device and check the mains
Motor off
voltage.
Heating off
- Switch off the device and check the mains
Motor off
voltage.
- Exit error state when any button is pressed.

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Lr 1000 control

Inhaltsverzeichnis