SW 8285-005: Tolerance adjustment (pH)
Adjustment of tolerance band. Select menu:
Parameter setting / Module pH/ORP / Cal preset values / Cal tolerance band
Menu Display
Menu selection
Return to meas
Parameter setting (Administrator)
System control
Module 8285 FRONT
Module 8285 BASE
Module 8285 El. COND
Module 8285 El. COND
Module 8285 pH/ORP
Return
Module 8285 pH/ORP (Administrator)
Input filter
Sensor data
Cal preset values
TC process medium
ORP/rH value
Delta function
Return
Cal tolerance band (Administrator)
Calibration data taken over when
tolerance band exceeded
Tolerance band check
Tolerance Zero
Tolerance Slope
Return
7.00 pH
17.1 °C
Select:
[enter]
Lingua
7.00 pH
17.1°C
7.00 pH
17.1 °C
Block
7.00 pH
17.1 °C
On
Off
+02.00 pH
002.0 mV/pH
Block
Parameter setting
Menu selection
From the measuring mode:
Press menu key to select menu.
Select parameter setting using arrow
keys, confirm with enter.
Passcode as delivered: 1989
Parameter setting
Select "Module pH/ORP",
confirm with enter.
pH/ORP module
Select Cal preset values using arrow
keys. Confirm with enter.
Then select Cal tolerance band.
Cal tolerance band
Enter tolerance ranges for zero and
slope
(enter numbers using arrow keys,
confirm with enter.)
15