TES-1351
CALIBRATION PROCEDURES
1. Using a acoustic calibrator
•
Make the following switch settings.
RANGE
RESPONSE
FUNCT
•
Insert the microphone carefully into the insertion hole
of the calibrator.
•
Tum on the switch of calibrator and adjust the
CAL screw of the instrument, until the level dis-
play indicates the desired level. Note: Our pro-
ducts
are
well
Recommended
calibrator
2. Calibration using the intemal oscillator
•
Make the following switch settings.
RANGE
RESPONSE
FUNCT
•
Display will show 94.0 ± 1.5dB
SOUND LEVEL METER
Hi
F
A
calibrated
before
shipment.
cycle
is
one
Hi
F
CAL 94dB
ACOUSTIC
CALIBRATOR
TE S
1351
SOUND LEVEL METER
dB
75. 0
RANGE
RESPONSE
FUNCT
Lo
S
A
Hi
C
F
POWER
MAX
CAL 94 dB
OFF
HOLD
Reset
Lo = 35
90dB
Hi = 75
130dB
IEC 651, ANSI S1.4 TYPE 2
ACCURACY: ± 1.5dB
VORGEHENSWEISE DER KALIBRIERUNG
1. Verwendung eines externen Kalibrierungsgerätes
•
Stellen Sie das Gerät wie folgt ein:
RANGE
RESPONSE
FUNCT
•
Fügen Sie das Mikrofon vorsichtig in die vorgesehene
Öffnung des Kalibriergerätes.
•
Schalten Sie den Kalibrator ein und drehen Sie die Ka-
librierungsschraube des Messgerätes bis im Display
der gewünschte Level eingestellt ist.
Hinweis: Unsere Produkte werden sorgfältig kalibriert,
year.
bevor sie unser Haus verlassen. Der empfohlene Kali-
brierungszyklus beträgt 1 Jahr.
2. Verwendung des internen Oszillators
•
Stellen Sie das Gerät wie folgt ein:
RANGE
RESPONSE
FUNCT
•
Das Display zeigt 94.0 ± 1.5dB
Hi
F
A
Hi
F
CAL 94dB
7